| Catalog Number |
RBYPCXL45 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the gastric varices using a packing coil xl, ruby xl, and a non-penumbra diagnostic catheter.While advancing a packing coil xl to the target location, the physician encountered resistance.Upon retraction, the embolization coil unintentionally detached within the diagnostic catheter.The physician subsequently used saline to flush the detached packing coil xl into the vessel.While advancing the a ruby xl, resistance was encountered again.The ruby xl was retracted and it unintentionally detached within the diagnostic catheter.The physician subsequently used saline to flush the detached ruby xl into the vessel.The procedure was completed using a new non-penumbra coil and the same diagnostic catheter.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|