| Model Number |
LUCAS 2 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device alarmed and stopped.In this state the device would not be able to provide compressions.If a device does malfunction, the device operating instructions indicate that the user is to perform manual chest compressions.There was no patient involvement reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
A stryker field service representative evaluated the device and duplicated the reported issue.The drive belt was replaced to resolve the issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing and the device was returned to the customer for service.
|
| |
|