Model Number IntelliVue Patient Monitor MX450
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
Philips received a complaint on a intellivue mx450 patient monitor indicating that the monitor displays a speaker malfunction error message.Additional information received from the customer confirmed that the device produced no audible sound.The device was in use on a patient at the time of event, no adverse event was reported.
 
Additional Manufacturer Narrative
E1 reporter phone #: (b)(6).Analysis was performed by remote service personnel which included troubleshooting.The reported issue was successfully replicated, and troubleshooting confirmed a speaker malfunction.Cross-check testing was conducted to support the diagnostic findings and conclusions.Based on the investigation, the reported problem was attributed to a speaker malfunction.Based on the results of the analysis, it was determined that the product has malfunctioned and the most likely cause of the reported problem was a speaker malfunction within the loudspeaker assembly.The customer was informed that the part needs to be replaced.A loudspeaker assembly ((b)(4), iv2-flex mechasy loudspeaker assembly), was ordered and shipped to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IntelliVue MX450 Patient Monitor
Common Device Name
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Manufacturer (Section D)
Philips Medizin Systeme Böblingen GmbH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
Philips Medizin Systeme Böblingen GmbH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
+49(703)14632031
MDR Report Key50852453
Report Number9610816-2026-101973
Device Sequence Number10053506
Product Code MHX
UDI-Device Identifier00884838038769
UDI-Public(01)00884838038769
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K130849
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIntelliVue Patient Monitor MX450
Device Catalogue Number866062
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/17/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-