The customer reported the connector tubing of the a-line insert kit being faulty and stated that, when inserting aline and connected tubing, they were unable to flush and saw what appeared to be an extra piece of plastic causing obstruction.A review of production records did not find any similar nonconformances noted by production or quality control personnel during first article, in-process, or case packaging inspections, and the vendor of the component reported in the complaint was notified.Considering the investigation results had not been received at that time, a root cause could not be positively identified.Due diligence was completed, no additional details were available, and there was no confirmed medical intervention or serious injury reported related to this incident.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|