| Catalog Number |
TRG400M |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "gloves are ripping.".
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported "gloves are ripping," and the investigation recorded the reported issue as "tears during use." the account provided a sample of three already torn gloves; there were no unused samples provided for additional testing, and thirteen gloves were selected from the provided samples and evaluated through donning and doffing.Medline confirmed the reported issue because one glove tore during this process, and a review of inspection records confirmed they met established quality requirements.Medline was unable to determine the root cause, and the quality team will continue to monitor for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|
| |
|