It was reported that the radiofrequency (rf) wire had difficulty to crossing the septum and advancing into the left atrium (la) and the patient experienced cardiac tamponade, pericardial effusion, and hypotension.During an ablation procedure to treat paroxysmal atrial fibrillation a versacross connect access solution for faradrive rf wire was used.A cavotricuspid isthmus ablation was performed with a non-boston scientific rf catheter prior to attempting the transseptal puncture (tsp).The versacross connect for faradrive system was positioned in optimal tsp position using fluoroscopy and intracardiac echocardiography (ice).The rf wire crossed the septum, but the tip got caught either on the septum or the la free wall during the one second rf cautery period.There was some difficulty advancing the wire and the physician stated this did not feel normal.The faradrive sheath was advanced into the la, and the versacross connect rf wire and dilator were removed from the patient.Preparations were made to insert a non-boston scientific mapping catheter into the sheath, but it was found that the sheath had lost transseptal access.Another tsp was to be attempted with the rf wire and dilator when it was found that the patient's blood pressure was 60/40 mmhg.The ice catheter was moved into position to observe the rv and an effusion was observed.Pericardiocentesis was performed, draining a minimal amount of blood.The procedure was cancelled due to the adverse event.The patient fully recovered and was discharged home.
|