MAUDE Adverse Event Report: MEDTRONIC, INC. Sphere-9; PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. Sphere-9; PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH Back to Search Results
Model Number AFR-00001
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a cardiac ablation procedure, during right atrial mapping, the catheter became stuck while crossing the tricuspid valve.The physician attempted to maneuver and disengage the catheter for several minutes without success, and it was suspected that it was caught in a papillary muscle.The procedure was temporarily interrupted.The catheter handle was cut off to exchange the short sheath for a longer sheath, which was used to facilitate catheter removal.Intracardiac echocardiography performed after removal showed that the tricuspid valve appeared to function normally, and no obvious tissue was observed on the catheter lattice.Access was regained, and the procedure continued with a new catheter without further incident.The case was completed. no patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sphere-9
Common Device Name
PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH
Manufacturer (Section D)
MEDTRONIC, INC.
880 technology park drive
billerica MA 01821
Manufacturer (Section G)
MEDTRONIC, INC.
880 technology park drive
billerica MA 01821
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
+353(091)708096
MDR Report Key50852815
Report Number3012520654-2026-00517
Device Sequence Number13149981
Product Code QZI
UDI-Device Identifier00763000993269
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P240013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAFR-00001
Device Catalogue NumberAFR-00001
Device Lot Number0234209699
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/22/2026
Date Device Manufactured07/16/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-