| Model Number |
AFR-00001 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a cardiac ablation procedure, during right atrial mapping, the catheter became stuck while crossing the tricuspid valve.The physician attempted to maneuver and disengage the catheter for several minutes without success, and it was suspected that it was caught in a papillary muscle.The procedure was temporarily interrupted.The catheter handle was cut off to exchange the short sheath for a longer sheath, which was used to facilitate catheter removal.Intracardiac echocardiography performed after removal showed that the tricuspid valve appeared to function normally, and no obvious tissue was observed on the catheter lattice.Access was regained, and the procedure continued with a new catheter without further incident.The case was completed. no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|