Block h6:
device code a020102 captures the reportable event of tip damaged.Block h11: pi analysis
upon receipt at our quality assurance laboratory, this access sheath underwent a thorough analysis.The device was visually and microscopically inspected.During the analysis, it was found that the tip was flattened and kinked, which was also confirmed when examining the device under magnification.Additionally, the sheath returned kinked.The device history record review was completed and confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.A review of the navigator hazard analysis was completed and confirmed that the events of device kinked, sheath bent/kinked and tip damaged were defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on the information available and analysis results, the kink identified in the sheath and the flattened tip could have caused by excess of force applied to the device such as operational factors during their use.Due to this, the events of sheath bent/kinked and tip damaged will be assigned a conclusion code of adverse event related to procedure.Based on the most relevant information of the complaint event description, device analysis and including the product record review results, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.Given this information, it is determined the investigation findings did not lead to a clear conclusion about the cause of the failure, all compiled information on this investigation determines that the investigation conclusion code is cause not established.
|