Catalog Number EL5ML
Medical Device Problem Codes Failure to Fire (2610); Component or Accessory Incompatibility (2897)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure the device does not work.Does not fit into the 5mm trocar.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/22/2026.D4 batch # a99v8l.Investigation summary.The product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was received with no apparent damage to the external components.In an attempt to replicate the reported incident, the device underwent functional testing.During evaluation, it was observed that clips failed to advance into the jaws upon subsequent trigger actuations.To further assess the condition of the internal components, the device was disassembled.Upon disassembly, the feed link was found to be broken, which disrupted the clip advancement mechanism and resulted in the observed feeding issue.Additionally, fourteen (14) clips were identified within the clip track.The broken feed link is considered the likely cause of the reported failure to feed clips into the jaws.The event reported was confirmed and it is related to improper use of the device.Possible causes for the damage found is due to the jaws may have been restricted during firing, or not fully through the trocar during firing.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch a99v8l number, and no non-conformances were identified.Additional information was requested, and the following was obtained: how did device not work? did device jammed (not fire clips)? did device not feed clips? did device drop or eject clips? did device sideways feed clips? did device fire malformed clips? did device fire scissored clips? if other please specify.I did not get a response from the surgeon.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
PR  
+1(303)552-6892
MDR Report Key50852973
Report Number3005075853-2026-07838
Device Sequence Number15958989
Product Code FZP
UDI-Device Identifier10705036001843
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K110699
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEL5ML
Device Lot NumberA99V8L
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/07/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/16/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1