Ethicon Endo-Surgery, LLC. LIGAMAX; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure the clip is stuck.Have there been reports of injuries from patients or users? no.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/22/2026.D4 batch # a98v1t.Investigation summary.The product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was received with no apparent damage.In addition, the tyvek packaging was returned along with the instrument.In an attempt to replicate the reported incident, the device underwent functional testing.During evaluation, the instrument was cycled; however, during subsequent actuations, the clips failed to advance into the jaws.To further assess the condition of the internal components, the device was disassembled.Upon disassembly, the feeder shoe tab was found to be damaged, which interfered with the clip advancement mechanism and resulted in the observed feeding issue.Additionally, ten (10) clips were identified within the clip track.Based on these findings, the damaged feeder shoe tab is considered the likely cause of the reported failure to feed clips into the jaws.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached regarding what may have caused the reported incident.A manufacturing record evaluation was performed for the finished device batch a98v1t number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
PR  
+1(303)552-6892
MDR Report Key50853024
Report Number3005075853-2026-07839
Device Sequence Number9936582
Product Code FZP
UDI-Device Identifier10705036001843
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K110699
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEL5ML
Device Lot NumberA98V1T
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/07/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/16/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-