(b)(4).Date sent: 9/22/2026.D4 batch # a98v1t.Investigation summary.The product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was received with no apparent damage.In addition, the tyvek packaging was returned along with the instrument.In an attempt to replicate the reported incident, the device underwent functional testing.During evaluation, the instrument was cycled; however, during subsequent actuations, the clips failed to advance into the jaws.To further assess the condition of the internal components, the device was disassembled.Upon disassembly, the feeder shoe tab was found to be damaged, which interfered with the clip advancement mechanism and resulted in the observed feeding issue.Additionally, ten (10) clips were identified within the clip track.Based on these findings, the damaged feeder shoe tab is considered the likely cause of the reported failure to feed clips into the jaws.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached regarding what may have caused the reported incident.A manufacturing record evaluation was performed for the finished device batch a98v1t number, and no non-conformances were identified.
|