A review of the complaint history for sn (b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6)was performed from the date of manufacture, 14-jun-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the device cubie tray was damaged.A field service engineer (fse) investigated the customer report that a cubie would pop out and not remain secured in position when replacing an old cubie with a new one in the row tray.The customer did not have a spare cubie available for testing, and an inspection of the cubie tray revealed no visible damage to its components.Based on the symptoms, a replacement cubie tray was determined to be necessary.Upon returning to the site with the required part, fse serviced the station and identified the first-row cubie tray in fh drawer 16 as damaged.Fse removed the defective cubie tray and installed a new replacement tray.Following the repair, the customer successfully inventoried the drawer pockets and verified proper operation, with all results confirmed as satisfactory.The system functioned as intended after the field service engineer repaired the device.
|