| Model Number |
EVIVA 0913-12T |
| Medical Device Problem Code |
Failure to Obtain Sample (2533)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported by the user site that the device was, "unable to lavage and get good tissue sampling.The problem was with the absorbing and cutting mechanism." customer outreach confirmed that "the issue was identified after the needle had been inserted into the patient" and "the needle was replaced with a new one, and the procedure was subsequently completed using the replacement needle".The "tissue specimen obtained was incomplete and consisted only of small tissue fragments.And could not be lavage." prior to changing the needle and successfully completing the procedure.Serious injury reporting is being submitted due to tissue not being acquired from the target requiring an additional needle insertion to complete the biopsy procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted for the reported lot number.The device was released meeting all qa specifications.Investigation is ongoing.
|
| |
|