The account reported finding defective filters, and it was reported that the circuit broke off on the inlet/outlet portion of the circuit where the filter is inside.A sample was received for investigation, a visual review of the sample determined that component 109889 cracked, finished good production records were reviewed, and the finished good pack build/bill of materials was reviewed.Based on the complaint details, sample evaluation, and investigation, the complaint is confirmed.The root cause for the reported issue cannot be determined; the filters already arrive connected to themselves by the manufacturer, and the only manipulation performed is to attach them to the tubing.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|