Catalog Number DYNJAA10685B
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "bh2 or general anesthesia circuit-lf- broken off at the end near the filter." "there were 2 circuits of this same lot that were found defective.".
 
Additional Manufacturer Narrative
The account reported finding defective filters, and it was reported that the circuit broke off on the inlet/outlet portion of the circuit where the filter is inside.A sample was received for investigation, a visual review of the sample determined that component 109889 cracked, finished good production records were reviewed, and the finished good pack build/bill of materials was reviewed.Based on the complaint details, sample evaluation, and investigation, the complaint is confirmed.The root cause for the reported issue cannot be determined; the filters already arrive connected to themselves by the manufacturer, and the only manipulation performed is to attach them to the tubing.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ANESTHESIA CIRCUIT-LF
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
madelyn ruppert
three lakes drive
northfield, IL 60093--275
MDR Report Key50853358
Report Number1417592-2026-47344
Device Sequence Number15866057
Product Code CAI
UDI-Device Identifier10198459372292
UDI-Public10198459372292
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJAA10685B
Device Lot Number25JBX325
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2025
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1