| Model Number |
P7900B |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Baxter received a report stating that centrella med-surg brake casters were not holding at the foot end of bed.The process step at which the issue occurred was not specified.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The baxter technician found the foot end casters needed to be replaced.Per the baxter service manual, it is necessary for the centrella bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Check the brakes to see whether the bed moves when the brakes are set.Replace parts or adjust the system if necessary.A search of baxter maintenance records showed baxter performed preventive maintenance on this bed in sep, 2025.It is unknown if the facility performed any other preventive maintenance on this bed.The technician replaced the foot end casters to resolve the reported event.Based on this information, no further action is required.
|
| |
|