The customer reported issue "too sticky" and "causing patients skin to peel off" with optifoam gentle ex, sacrum, 9x9; according to the facility, "when the dressing is removed on day 3 the dressings are extra sticky and removing patients¿ skin," the dressing was applied on "very fragile skin," and the skin was described as "red and inflamed." the received sample was evaluated, there was nothing abnormal observed with the sample¿s adhesion, and the supplier tested adhesion and determined the peel adhesion of the return sample was within the product specification.The reported issue could not be confirmed with the received sample and additional answers given by the sales rep.No root cause was identified; no medical intervention, serious injury, or follow-up care has been reported, no adverse trending was identified, and the complaint was closed.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|