• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; SHEET,DRAPE,40X58,STERILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; SHEET,DRAPE,40X58,STERILE Back to Search Results
Catalog Number DYNJP2410
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "we have experienced sterile field contamination due to fluids penetrating through the (b)(6) half-drape.".
 
Additional Manufacturer Narrative
The customer reported that the (b)(6) does not work for their needs and stated that the drapes were "not completely sealed/impermeable" and dye "passed through onto the table underneath the drape." the customer was contacted and requested to provide additional information; samples were reviewed, 14 each of the (b)(6) were received, the samples were measured and met product specifications, and the (b)(6) has an aami level 1 rating.Based on the results of the investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.This is a customer preference issue and needs a high-level aami rated drape, and no trend was identified.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
SHEET,DRAPE,40X58,STERILE
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
azmat shah
three lakes drive
northfield, IL 60093--275
MDR Report Key50853804
Report Number1417592-2026-47390
Device Sequence Number9936607
Product Code PUI
UDI-Device Identifier10080196739691
UDI-Public10080196739691
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJP2410
Device Lot Number11025080457
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/17/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-