The customer reported that the (b)(6) does not work for their needs and stated that the drapes were "not completely sealed/impermeable" and dye "passed through onto the table underneath the drape." the customer was contacted and requested to provide additional information; samples were reviewed, 14 each of the (b)(6) were received, the samples were measured and met product specifications, and the (b)(6) has an aami level 1 rating.Based on the results of the investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.This is a customer preference issue and needs a high-level aami rated drape, and no trend was identified.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|