| Model Number |
IntelliVue X3 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on an intellivue x3 patient monitor indicating that the ecg waveform was displayed clearly, but the heartrate (hr) count was doubled.Replacing the electrode patch did not resolve the issue.When other devices were used on the same patient, the hr was correct.The device was in use on a patient at time of event, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed by onsite service personnel which included operational checks using a simulator.The results of the analysis were unable to reproduce the issue.Based on the results of the analysis, the product was confirmed to be operating per specifications, and the cause of the reported problem was unable to be determined.The device was returned to use, and the customer was advised to monitor and device and advise if the issue occurs again.A review of the service history for this device shows the problem has not been reported again for this device.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.
|
| |
|