The customer stated that the pendant had broken buttons, and that the head section would go up but did not go down all the way, while the foot section worked fine.Medline received one bed "pendant only" in used condition with wear marks on the buttons and power cord damage, and bench tested it with known working components; the head down button did not work and all other buttons did work.The investigation results were confirmed and the customer complaint was confirmed.The cause was documented as cannot determine cause with received sample; the issue may be due to an inner electrical issue, and the division will monitor through trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|