• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Olympus Winter & Ibe GmbH Berlin Facility HF unit "ESG-400"; Electrosurgical generator ESG-400

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Olympus Winter & Ibe GmbH Berlin Facility HF unit "ESG-400"; Electrosurgical generator ESG-400 Back to Search Results
Model Number WB91051W
Medical Device Problem Codes Unintended Power Up (1162); Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported after switching on system it automatically restarted with e433 error during set up.There was no patient injury reported.During the initial evaluation it was observed that a defective generator board caused error e433 and automatic restarted the device.
 
Additional Manufacturer Narrative
Date of event is unknown.Additional information will be provided if available.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF unit "ESG-400"
Common Device Name
Electrosurgical generator ESG-400
Manufacturer (Section D)
Olympus Winter & Ibe GmbH Berlin Facility
rheinstrasse 8
teltow, brandenburg 14513
GM  14513
Manufacturer (Section G)
Olympus Winter & Ibe GmbH Berlin Facility
rheinstrasse 8
teltow, brandenburg 14513
GM   14513
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50854064
Report Number3003724334-2026-00499
Device Sequence Number15936832
Product Code GEI
UDI-Device Identifier04042761076838
UDI-Public04042761076838
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K203682
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWB91051W
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/08/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/15/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-