• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline ReNewal MEDLINE; LIGASURE MARYLAND 37CM FT10 COMPATIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline ReNewal MEDLINE; LIGASURE MARYLAND 37CM FT10 COMPATIBLE Back to Search Results
Catalog Number LF1937CRH
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "during surgery the device failed in the way of the instrument jaw separating from the housing during use.".
 
Additional Manufacturer Narrative
The customer reported that during surgery a ligasure maryland jaw laparoscopic device failed in the way of the instrument jaw separating from the housing during use.During the investigation, the received device was inspected and tested for nonconformities per internal procedure; clinical debris was observed on the jaws and handle of the device, and the jaw had separated from the device.Reported issue confirmed was documented as n/a - oem, and the exact root cause of the reported issue could not be determined.Due to the device having an oem paddle, oem packaging, and oem lot number, and no observable medline renewal reprocessing marks, it was determined that the received device was oem and had not been reprocessed by medline renewal.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
LIGASURE MARYLAND 37CM FT10 COMPATIBLE
Manufacturer (Section D)
Medline ReNewal
1500 ne hemlock avenue
redmond OR 94456 -187
Manufacturer Contact
brianna montgomery
three lakes drive
northfield, IL 60093
MDR Report Key50854105
Report Number3032391-2026-00137
Device Sequence Number9553926
Product Code NUJ
UDI-Device Identifier10888277719415
UDI-Public10888277719415
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberLF1937CRH
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/27/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2023
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Patient SexUnknown
-
-