According to the facility, during a procedure the ligasure "was sticking when used at the beginning of the procedure"; the item was replaced and the procedure was completed without injury or medical intervention.During the investigation, the device was inspected and tested for nonconformities per internal procedure; clinical debris was noted in the jaws and on the handle of the device, no abnormalities were observed during the rfid data review, and simulated use was performed using beef tripe test media.Upon connection, the device was recognized by the console and no error messages appeared; during simulated use, the device sealed and cut the test media as expected, and no mechanical/functional issues were observed with the device.The reported issue was not confirmed, and the exact root cause of the reported issue could not be determined, as the reported issue could not be replicated.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|