| Catalog Number |
MDS88200HL |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that ¿the hoyer lift is leaking fluid".
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that ¿the hoyer lift is leaking fluid.Upon technician diagnosis it was found that hydraulic fluid was indeed leaking from the hydraulic cylinder.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.
|
| |
|