The complaint reported ¿components are broken or mishaped within the packs, the packs show no signs of damage on the shelf or when being opened¿ and identified defects with ¿bd 10 ml syringe (found in 2 seperate packs) pitcher (found 3 seperate packs) basin¿; clarification stated ¿during set up bent emesis basin,¿ ¿during surgery syringe broke,¿ ¿during set up, syringe was broken and damaged pitcher,¿ and ¿cracked pitcher,¿ with ¿did an injury occur? :no.¿ we received three photos and no physical sample was available; finished good production records and finished good pack build were reviewed, with no line down situations, no run without approvals, deviations or non-conformance reports that might have had an impact on this issue, and the component placements did not contribute to the reported damage.Upon reviewing the photographs, the emesis basin was slightly bent on the exterior surface, confirming the reported issue; the cracked pitcher defect was unable to be confirmed, as no physical sample or detailed photographic evidence was provided to support the investigation.For the bent emesis, the item was either damaged prior to placement in the pack and was not noted during assembly or the item became damaged due to rough handling of the pack; the division will monitor this issue for trending purposes, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|