The account reported finding wet packs for the dbd-arthroscopy pack, and the complaint stated that wet packs were potentially used on patient.No sample or photo was received for the investigation; finished good production records were reviewed and no non-conformances were noted that would have contributed to the reported issue, the finished good pack build was reviewed, trending was reviewed, and a stock check found no inventory left of this remaining lot at the branch.The reported problem could not be evaluated with a sample or photo; based on the complaint details, medline determined that the reported damage occurred after kit assembly during transit or storage of the pack, and the root cause is considered an operations service issue.Per the facility, the device was not used on a patient and therefore there was no patient involvement; there was no serious injury identified, follow up care needed or medical intervention reported related to the incident, and the division will monitor for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|