The account reported finding ra separated from manifold, and the complaint stated that the issue occurred during use and no injury occurred.A photo was received to aid the investigation; no physical sample was available, finished good production records were reviewed, and finished good pack build records were reviewed with no non-conformances noted that would have contributed to the reported issue.Based on the information provided and review of the photo received from the customer, the report of an ra separated from the manifold can be confirmed.However, without the physical sample available for review, it is impossible to determine if there was a successful swage of the ra to the manifold stem, therefore, the root cause could not be identified; the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|