• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Synthes GmbH VOLT; Screw, fixation, bone

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Synthes GmbH VOLT; Screw, fixation, bone Back to Search Results
Catalog Number 02.535.130
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that volt cort scr ø3.5 t15 l32 sst and volt cort scr ø3.5 t15 l30 sst were involved in a revision nonunion femur case with another company¿s reconstruction nail.After insertion of the nail, the femur was provisionally fixated to compress the nonunion.During insertion of the two 3.5 cortex screws, one in the plate and one for the articulating reduction device, the screw heads sheared off.Both broken screws remained embedded in the patient, with one retained in the plate and the other breaking during shaft extraction.The plate remained properly fixated, and no patient harm or secondary impact related to the embedded devices was reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional manufacturer narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLT
Common Device Name
Screw, fixation, bone
Manufacturer (Section D)
Synthes GmbH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
Monument
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
+1(908)386-3534
MDR Report Key50854270
Report Number8030965-2026-09630
Device Sequence Number13275196
Product Code HWC
UDI-Device Identifier10886982324467
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K233665
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number02.535.130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-