A complaint was received regarding cracked gastrostomy tubes; per the customer, the reported issue was discovered prior to use and has happened on two separate occasions.A sample was provided by the customer, the vendor was notified of the issue and sent the form along with the sample received, and the vendor¿s review of the production processes and relevant records noted that there were no nonconformities during the manufacture of this product lot; the vendor noted that the returned sample showed obvious signs of use as it was filled with fluid and the crack was noted to be a jagged tear rather than a clean cut, pointing to possible user error.Based on the information provided and the results of the investigation, the reported issue was confirmed; however, the vendor was unable to determine that the issue was related to the manufacture of the product, and neither a root cause nor a systematic trend could be identified.No corrective actions have been warranted at this time, and a divisional quality representative will continue to monitor this issue for tracking and trending purposes during quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|