MAUDE Adverse Event Report: Synthes GmbH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 18MM; Screw, fixation, bone

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Synthes GmbH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 18MM; Screw, fixation, bone Back to Search Results
Catalog Number 04.211.018TS
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2026, the patient underwent orif for a clavicle shaft fracture using the screw in question.The vacp was used for fixation.During screw insertion, excessive torque was continuously applied to the screw, resulting in screw fracture.The proximal one-third of the screw fractured, while the remaining distal two-thirds remained implanted in the patient.The surgery was completed successfully with the retained fragment left in the patient.According to the surgeon, the risk of screw fracture increased because the screw size had been reduced to 2.7 mm compared to the previous product.The surgery was completed successfully without any surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 18MM
Common Device Name
Screw, fixation, bone
Manufacturer (Section D)
Synthes GmbH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
Synthes Selzach
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
+1(908)386-3534
MDR Report Key50854670
Report Number8030965-2026-09633
Device Sequence Number9652009
Product Code HWC
UDI-Device Identifier10886982119018
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103243
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.211.018TS
Device Lot Number462L209
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/19/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age24 YR
Patient SexMale
-
-