Synthes GmbH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 18MM; Screw, fixation, bone
|
Back to Search Results |
|
| Catalog Number |
04.211.018TS |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, the patient underwent orif for a clavicle shaft fracture using the screw in question.The vacp was used for fixation.During screw insertion, excessive torque was continuously applied to the screw, resulting in screw fracture.The proximal one-third of the screw fractured, while the remaining distal two-thirds remained implanted in the patient.The surgery was completed successfully with the retained fragment left in the patient.According to the surgeon, the risk of screw fracture increased because the screw size had been reduced to 2.7 mm compared to the previous product.The surgery was completed successfully without any surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|