| Catalog Number |
MP1000 CHINA |
| Medical Device Problem Codes |
Crack (1135); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation result:\
a mp1000 china product was not available for investigation; however, the customer confirmed that the complaint sample was from lot 25085182.The feedback provided by the customer states that, \"a crack formed at the connection between the needle-free connector and the infusion set during an iv infusion, causing fluid leakage\".No further information was available to assist the investigation in this instance.\
\
the details of this feedback were forwarded to the manufacturing site for investigation, where a review of the production records for lot 25085182 did not identify any in-process testing failures or quality deviations which may have caused or contributed to a report of this nature.\
\
the root cause of the customer¿s experience could not be determined in this instance as no sample was received for investigation.Without a sample to examine it is not possible to determine whether a manufacturing defect could have caused or contributed to a report of this nature.\
\
in this instance, without the complaint sample to examine it has not been possible to conclusively link this feedback to a specific failure mode; however, a review of the customer feedback database indicates that this is a rare occurrence with a small number of similar reports against the mp1000 china products in the past 12 months.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Event or Problem Description
|
|
It was reported, maxplus, crack, causing leak.\
\
the following was provided by the initial reporter: a crack formed at the connection between the needle-free connector and the infusion set during an iv infusion, causing fluid leakage.
|
| |
|
Search Alerts/Recalls
|