Customer reported that during aline preparation for patient, the tubing leaked at one of the connections when flushed and saline leakage at connection was recognized.A review of production records did not find any similar nonconformances noted by production or quality control personnel during first article, in-process, or case packaging inspections, and the vendor of the component reported in the complaint was notified.There was no sample available for this complaint, and the lack of an available sample prevented confirmation of the reported event.Considering the investigation results, a root cause cannot be positively identified; the division will continue to monitor this issue and will initiate corrective actions if a trend is identified.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|