The account reported that the pressure infuser bags were not holding pressure.A sample was received; the bag was inspected for any defects and none were found, filled to 300mmhg and closed off with the attached clamp, compressed with no signs of pressure loss, and then deflated with no issue; this cycle was repeated two more times with the same result.The reported issue was not confirmed, and the cause was documented as not confirmed with received sample.Our quality team will monitor this issue for trending purposes, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|