The customer reported mold found in sterile water and later stated that, after looking at the product, "it has been discovered that it is not mold, it is a string (foreign object)." medline quality visually inspected the samples and found 30 were fully sealed and one sample was open; there appears to be a string, yellow/green in color and approximately 1" in length, floating inside the bottle, and all remaining bottles appeared unaffected.As a result, medline confirmed the reported complaint.Per the vendor, the root cause of the string in the bottle could possibly be from a production technician's clothing; the division will monitor the complaint during quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|