It was reported that a patient with an artificial urinary sphincter (aus) complained about the functioning of the device.The physician ordered imaging, which confirmed that the balloon contained 17 cc of fluid.This could possibly indicate a fluid leak; however, this was not confirmed by the physician.The physician added 5 cc of saline and will reassess the patient to determine whether a complete revision is necessary.No additional information was provided.
|
|
Detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review was unable to be performed as the lot number was not provided.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of leaks, not specified was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
|