Medline Industries, LP MEDLINE; DBD-ROBOTIC PACK Back to Search Results
Catalog Number DYNJ56511C
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "every pack they have contains a faulty 22g safety needle.The safety flops down when they are trying to draw up.".
 
Additional Manufacturer Narrative
The account reported finding a defective safety needle, and cookeville clinical staff said the safety mechanism flops around when they are drawing up into the syringe.A sample was received; medline received a bag full of needles still sealed inside of its vendor packaging, observed no visual defects or damage to the component, performed functional testing, reviewed the pack's production record, and reviewed the finished good pack build.Functional testing of the component did exhibit the safety dropping; however, this did not affect use of the item and is the intended functionality, and the component was made to specification.Based on this, the root cause is considered customer preference, and the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-ROBOTIC PACK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
venkata peruri
three lakes drive
northfield, IL 60093--275
MDR Report Key50854851
Report Number1417592-2026-47602
Device Sequence Number15936719
Product Code LRO
UDI-Device Identifier40195327326280
UDI-Public40195327326280
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ56511C
Device Lot Number23EBC738
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/10/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2023
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-