Medline quality received a complaint that tubing attached primary line to catheter and not separate and luer lock was damaged.The complaint was re-opened due to receiving the sample; medline evaluated the tubing that was returned and did not observe any physical deformities or any damage.The reported concern was not confirmed with the received sample, and no root cause was identified.A retrospective review of the past year indicated that there have been no prior complaints in regard to this issue, and the division will continue to monitor this issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|