The reported issue was "item not closing", and the complaint transcript stated, "item wasn't working.They weren't closing".Medline quality received 1 cs of the item for evaluation; visual inspection of the provided product revealed no defects nor abnormalities, and functional testing was conducted by connecting the product to a vacuum and observing whether it suctioned liquid as intended.Functional testing of the product revealed no defects nor abnormalities, and the reported concern could not be confirmed, as the sample functioned normally; the cause was documented as "not confirmed with received sample".The division will continue to monitor this issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|