• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBD-OPEN HEART PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBD-OPEN HEART PACK Back to Search Results
Catalog Number DYNJ84416
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "it looks like the heat weld failed and the sms separated from the plastic.The other complaint is that the adhesive on the chest incision incise window does not stick to the skin well.".
 
Additional Manufacturer Narrative
The account reported that the cv incise drape failed during surgery, the sms separated from the plastic, and the adhesive on the chest incision incise window does not stick to the skin well.A photo and a sample were received for investigation; the pictures show separation between the grey pe material and the base material, the sample was reviewed and there was separation between the two materials, finished good production records had no non-conformances, and there were no issues with the pack build.The separation between the two materials was confirmed, while the adhesive issue with the drape was reviewed and the adhesive was found to meet the product specifications.Based on the complaint details the root cause is considered to be a vendor product issue relating to the component manufacturing process; it was reported that there was no impact to the patient, staff or procedure, and no medical intervention, serious injury, or follow-up care has been reported.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-OPEN HEART PACK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
urmi parekh
three lakes drive
northfield, IL 60093--275
MDR Report Key50855260
Report Number1417592-2026-47672
Device Sequence Number13250140
Product Code OEZ
UDI-Device Identifier40195327363292
UDI-Public40195327363292
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ84416
Device Lot Number23EBM662
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2023
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1