The account reported that the cv incise drape failed during surgery, the sms separated from the plastic, and the adhesive on the chest incision incise window does not stick to the skin well.A photo and a sample were received for investigation; the pictures show separation between the grey pe material and the base material, the sample was reviewed and there was separation between the two materials, finished good production records had no non-conformances, and there were no issues with the pack build.The separation between the two materials was confirmed, while the adhesive issue with the drape was reviewed and the adhesive was found to meet the product specifications.Based on the complaint details the root cause is considered to be a vendor product issue relating to the component manufacturing process; it was reported that there was no impact to the patient, staff or procedure, and no medical intervention, serious injury, or follow-up care has been reported.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|