This is a combination product
the customer reported dressing not adhering well on patients even when skin is dry on dressing application, dressing not staying in place, and redness/skin reaction with observation of redness under dressing on some patients; 2 staff applied dressings and noted skin irritation (redness only, no itching) after removal.The investigation evaluated a received sample, and the supplier reviewed batch records and confirmed that the item had been manufactured without any issues; the product has undergone biocompatibility testing per iso 10993 and has passed.The customer reported issue "redness/skin reaction" could not be confirmed with the received sample, and the product did not show any abnormalities.No root cause was identified; no adverse trending identified, and the complaint was ok to close.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|