The account reported finding device cut the under skin; it was also reported that when the physician clamps down on the foreskin, it cuts the skin by just closing the clamp, on the underside of the penis.A sample was received for the investigation; functional inspection was conducted with the provided sample, and it was found that the item function properly, and simulated grip strength testing with a latex glove found the glove was not cut or teared.The opening was measured as 2.3mm, which meets all the requirements and specifications, and based on the complaint details, sample evaluation and investigation, the complaint cannot be confirmed.The cause was not confirmed with received sample, and the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|