| Model Number |
2110 |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was found during the investigation of a returned pump that an error "41622" was pops up as soon as the pump turns on due to faulty flex sensor.This failure condition could potentially interrupt therapy and may stop the pump if it occurs during patient use.There was no patient involvement or harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation.Visual inspection identified a missing battery door.The device event log and 12-month service history were reviewed.Functional testing confirmed error code 41622 upon pump power-up.Replacement of the flex sensor resolved the reported issue.During further testing, error code 46440 was observed and resolved by replacing the dso sensor.The root cause of the reported 41622 error was determined to be a faulty flex sensor.
|
| |
|