|
The customer reported that the item caused skin irritation and that it rips because the adhesive is too sticky; the skin looked red and irritated, with no treatment, serious injury, medical intervention, follow-up care, or procedure impact reported.Medline quality received one sample for evaluation, and physical inspection was conducted with the provided sample.The customer did not provide any evidence of skin irritation caused by the grounding pad and at this time this issue cannot be confirmed; at the same time the issue with the plug has been confirmed, as the plug was damaged.The suspected root cause is a manufacturing issue, and the division will continue to monitor the issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|