|
The customer advised that there is a kink in the tubing and cannot be used.Medline quality received 1 ea for evaluation, and visual inspection was conducted with the provided sample; it was found that a kink was in the tubing.At this time, the reported concern is confirmed, as kink was found, and the suspected root cause is undetermined at this time.The manufacturing site has been notified of this issue, and the division will continue to monitor the issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|