This vmsr submission is part of a retrospective data remediation review.The included event represents one malfunction event that medline became aware of during q1 2025.Associated lot numbers: 24aoa011 and 24joa067.Associated di: 10197344012367.No serial number was available.One malfunction event involving 24 device is summarized in this report.One device was returned to the manufacturer and evaluated.The returned device was labeled for single use.No devices were identified as reprocessed or reused.The account reported that the medline unheated adult breathing circuit, catalog number hud1613, was not passing leak tests at multiple facilities.The device was reviewed at the medline nuevo laredo manufacturing facility.Visual inspection found that the original packaging had previously been opened, and functional testing confirmed leakage in the returned sample.Device history records for lots 24aoa011 and 24joa067 were reviewed; no related issues, discrepancies, or associated nonconformance reports were identified, and the records showed that the product was assembled and inspected according to specifications.The investigation found insufficient evidence that the issue originated during manufacturing assembly or molding.Because the sample was not received in its original packaging and its condition limited the evaluation, the root cause could not be identified and no corrective action could be established.The complaint was provided to the component manufacturing facility for further evaluation, the manufacturer/vendor investigation was received, and the quality team will monitor the issue for trending purposes.Trending identified no additional complaints for this issue for the account and four additional complaints for this product and issue during the prior 24 months.The complaint review was completed, and the reportability determination did not change.The complaint was later reopened under capa-02381 for retrospective remediation review; that review identified no new information, failure modes, trends, or additional patient-safety or product-performance risks requiring revision of the complaint classification or further investigation.Complaints remain subject to monitoring under sop-00032, with appropriate action to be taken if an objectionable trend is identified.No patient information was available.There were no reports of death, serious injury, medical intervention, or follow-up care.It has been determined that the reported event could cause or contribute to serious injury if it were to recur.In an abundance of caution, this is a reportable malfunction event.If a sample, photograph, or additional information becomes available, the complaint and report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|