| Model Number |
WA50040A |
| Medical Device Problem Codes |
Crack (1135); Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the rigid video laparoscope to the service center for a separate issue.During the device analysis, the service repair technician indicated that a light guide lens with a crack and a noisy image were found.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation regarding the light guide lens with a crack, the cause was a component failure of the distal end cover glass.For the noisy image, the cause was a damaged charge coupled device unit.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|