It was reported that the patient underwent a gynecare tvt implant procedure and had a favorable outcome of spontaneous urination without difficulty or recurrence of incontinence.However, for a few months, the patient experienced the reappearance of a more marked emergency, associated with malodorous vaginal discharge.Endoscopic exploration was therefore programmed to eliminate erosion of the strip.In (b)(6) 2022, the patient had a clinical examination and the presence of anterior vaginal erosion, facing the urethra measuring approximately 1 cm wide by 4 mm high, with a strip that was visible.In (b)(6) 2025, the patient had a cystoscopy and vaginoscopy.It was found that the patient had an absence of involvement of urethra and bladder and an absence of other areas of erosion or associated abnormality.It was decided to proceed to an excision of the exposed strip fragment, associated with vaginal covering.The patient was informed of the risk of recurrence of urinary incontinence following this procedure.It was also noted that the patient has multiple sclerosis and is currently taking ocrevus.In addition, the patient is autonomous in walking, has a known neuro-bladder, and mixed urinary incontinence with also some imperiosities in the context of her neurovessie.
|
|
The neuchatel team received one gynecare tvt device 810041bl, lot number: 3935674 (part of mesh) for device analysis.An investigation was performed on the received product and on the batch record file.The received device was manipulated and explanted as original packaging and all components were missing.Only the remaining part of the mesh was visible with lot of organic matter around it.Based on the evaluation, this complaint is not related to a manufacturing problem.The defect observed during the evaluation corresponds to the description of the event (vaginal erosion) and cannot be confirmed with the received product.The product was conforming to specifications at the release.A manufacturing record evaluation was performed for the finished lot number 3935674 (product code 810041bl) and no non-conformances or manufacturing irregularities were identified.Expiration date: 31oct2022 manufacturing date: 26nov2019.
|