| Model Number |
BF-XP190 |
| Medical Device Problem Codes |
Erratic or Intermittent Display (1182); Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the bronchovideoscope exhibited an image issue (image flickers during angulation).The malfunction was identified during reprocessing.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: image flickers when angulating due to damage to the charged coupled device.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.Date of event is unknown.Additional information will be provided if available.
|
| |
|