Section a patient information: refer to section b5 for complete patient information.E1 postal code is not available, therefore 00000 was entered for form requirement.E1 phone number reached maximum character limit, the complete phone is (b)(6).An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
|
During the requested site visit, the field service engineer (fse) observed that nozzle #5 was clogged and replaced the nozzle c4 #1, #4, #5 (rohs) (7-205228-02).The fsr also replaced additional components, including the worn tubing, peristaltic head (rohs) (7-202464-01) and peristaltic head (rohs) (7-202464-01) which resolved the discrepant result issue.A review of the alinity c processing module, serial number (b)(6) service history found no contributing factors on or around the date of the complaint.A review of tracking and trending of the alinity c processing module or parts did not identify any trends associated with the complaint issue.A review of historical data revealed no trends, systemic issues, or related nonconformances.The alinity ci-series operations manual addresses operational precautions and limitations and troubleshooting of erratic/discrepant results.The alinity c service documentation provides component replacement, removal and verification of the replacement of parts.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.Based on the investigation, no product deficiency was identified for either the alinity c processing module, serial number (b)(6), or the nozzle c4 #1, #4, #5 and peristaltic head (rohs).
|