The account reported finding assembly issue; the customer stated that the product was received with tubing improperly positioned and protruding from the side of the pack, compromising sterility.No sample or photo was received for investigation; finished good production records and finished good pack build were reviewed, and there were no line down situations, deviations, run-without, or non-conformance reports that might have had an impact on this issue.At this time, the reported defect cannot be confirmed because no physical sample or detailed photographic evidence was provided; the incident cannot be confirmed due to the lack of photographic or physical evidence.The division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|