| Model Number |
72404131 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
05/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with this artificial urinary sphincter (aus) experienced recurring urinary incontinence.A surgical procedure was performed to fully replace the aus.It was determined that the recurring urinary incontinence was associated with the device length selected and device fatigue over time.No additional patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant product implanted at the time of the event:
model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 0155229013.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 101371005.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).
|
| |
|
Search Alerts/Recalls
|