| Model Number |
MAJ-210 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, a piece of the biopsy valve rubber fell off into the patient¿s body but was retrieved immediately.It is unknown which part of the body the rubber piece fell into and what tool was used for retrieval.The procedure was completed after switching to another biopsy valve.There were no further reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 - initial reporter establishment name : (b)(6).The device was returned to olympus for inspection, and the reported failure was confirmed.It was found that part of the biopsy valve was broken, and the shape of the detached rubber fragment matched the damaged portion of the biopsy valve.Based on the results of the investigation, a root cause could not be identified.It is possible that the biopsy valve broke due to stress caused by inserting or removing the syringe at an angle.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|